Potential advantages of the forced oscillation technique over other methods for monitoring total respiratory mechanics during artificial ventilation are that it does not require patient relaxation, and that additional information may be derived from the frequency dependence of the real (Re) and imaginary (Im) parts of respiratory impedance. We wanted to assess feasibility and usefulness of the forced oscillation technique in this setting and therefore used the approach in 17 intubated patients, mechanically ventilated for acute respiratory failure. Sinusoidal pressure oscillations at 5, 10 and 20 Hz were applied at the airway opening, using a specially devised loudspeaker-type generator placed in parallel with the ventilator. Real and imaginary parts were corrected for the flow-dependent impedance of the endotracheal tube; they usually exhibited large variations during the respiratory cycle, and were computed separately for the inspiratory and expiratory phases. In many instances the real part was larger during inspiration, probably due to the larger respiratory flow, and decreased with increasing frequency. The imaginary part of respiratory impedance usually increased with increasing frequency during expiration, as expected for a predominately elastic system, but often varied little, or even decreased, with increasing frequency during inspiration. In most patients, the data were inconsistent with the usual resistance-inertance-compliance model. A much better fit was obtained with a model featuring central airways and a peripheral pathway in parallel with bronchial compliance. The results obtained with the latter model suggest that dynamic airway compression occurred during passive expiration in a number of patients. We conclude that the use of forced oscillation is relatively easy to implement during mechanical ventilation, that it allows the study of respiratory mechanics at various points in the respiratory cycle, and may help in detecting expiratory flow limitation.
This article has been cited by other articles:
![]() |
M. K. Johnson, M. Birch, R. Carter, J. Kinsella, and R. D. Stevenson Use of reactance to estimate transpulmonary resistance Eur. Respir. J., June 1, 2005; 25(6): 1061 - 1069. [Abstract] [Full Text] [PDF] |
||||
![]() |
R. Farre and D. Navajas Assessment of expiratory flow limitation in chronic obstructive pulmonary disease: a new approach Eur. Respir. J., February 1, 2004; 23(2): 187 - 188. [Full Text] [PDF] |
||||
![]() |
R.L. Dellaca, P. Santus, A. Aliverti, N. Stevenson, S. Centanni, P.T. Macklem, A. Pedotti, and P.M.A. Calverley Detection of expiratory flow limitation in COPD using the forced oscillation technique Eur. Respir. J., February 1, 2004; 23(2): 232 - 240. [Abstract] [Full Text] [PDF] |
||||
![]() |
E. Oostveen, D. MacLeod, H. Lorino, R. Farre, Z. Hantos, K. Desager, and F. Marchal The forced oscillation technique in clinical practice: methodology, recommendations and future developments Eur. Respir. J., December 1, 2003; 22(6): 1026 - 1041. [Abstract] [Full Text] [PDF] |
||||
![]() |
M. YOUNES, J. KUN, B. MASIOWSKI, K. WEBSTER, and D. ROBERTS A Method for Noninvasive Determination of Inspiratory Resistance during Proportional Assist Ventilation Am. J. Respir. Crit. Care Med., March 15, 2001; 163(4): 829 - 839. [Abstract] [Full Text] |
||||
![]() |
A.M. Lorino, F. Lofaso, E. Dahan, A. Harf, and H. Lorino Respiratory impedance response to continuous negative airway pressure in awake controls and OSAS Eur. Respir. J., January 1, 2001; 17(1): 71 - 78. [Abstract] [Full Text] [PDF] |
||||
![]() |
W. J. Randerath, K. Parys, F. Feldmeyer, B. Sanner, and K. H. Ruhle Self-adjusting Nasal Continuous Positive Airway Pressure Therapy Based on Measurement of Impedance: A Comparison of Two Different Maximum Pressure Levels Chest, October 1, 1999; 116(4): 991 - 999. [Abstract] [Full Text] [PDF] |
||||
![]() |
A. M. Lorino, K. Hamoudi, F. Lofaso, E. Dahan, C. Mariette, A. Harf, and H. Lorino Effects of continuous negative airway pressure on lung volume and respiratory resistance J Appl Physiol, August 1, 1999; 87(2): 605 - 610. [Abstract] [Full Text] [PDF] |
||||
![]() |
A. M. LORINO, F. LOFASO, D. DUIZABO, F. ZERAH, F. GOLDENBERG, M. P. d'ORTHO, A. HARF, and H. LORINO Respiratory Resistive Impedance as an Index of Airway Obstruction during Nasal Continuous Positive Airway Pressure Titration Am. J. Respir. Crit. Care Med., November 1, 1998; 158(5): 1465 - 1470. [Abstract] [Full Text] [PDF] |
||||