Original Articles |
Unlike conventional methods, the interrupter method for measuring airway resistance is non-invasive and requires minimal patient co-operation. It can therefore be applied in critically ill patients, acute asthmatics, neonates, pre-school children, geriatric patients and unconscious patients. The method is based on transient interruption of airflow at the mouth for a brief period during which alveolar pressure equilibrates with mouth pressure. Measurement of mouth pressure is used to estimate alveolar pressure prior to interruption and the ratio of this to flow prior to interruption gives airway resistance. Using the interrupter method we have developed a portable device for measuring airway resistance which is simple to use and gives a direct instantaneous reading. Measurements of airway resistance obtained using the new device were compared with those obtained using conventional body plethysmograph methods in 43 adult patients. A close correlation was seen (r = 0.86). The two methods appear equally sensitive in detecting changes in airway resistance following bronchodilator therapy. The device has been used successfully in pre-school children unable to co-operate with conventional methods.
This article has been cited by other articles:
![]() |
N. Beydon, S. D. Davis, E. Lombardi, J. L. Allen, H. G. M. Arets, P. Aurora, H. Bisgaard, G. M. Davis, F. M. Ducharme, H. Eigen, et al. An Official American Thoracic Society/European Respiratory Society Statement: Pulmonary Function Testing in Preschool Children Am. J. Respir. Crit. Care Med., June 15, 2007; 175(12): 1304 - 1345. [Full Text] [PDF] |
||||
![]() |
Y. F. Guo, F. Herrmann, J-P. Michel, and J-P. Janssens Normal values for respiratory resistance using forced oscillation in subjects >65 years old Eur. Respir. J., October 1, 2005; 26(4): 602 - 608. [Abstract] [Full Text] [PDF] |
||||
![]() |
P.J.F.M. Merkus, H.G.M. Arets, T. Joosten, A. Siero, M. Brouha, J.Y. Mijnsbergen, J.C. de Jongste, and C.K. van der Ent Measurements of interrupter resistance: reference values for children 3-13 yrs of age Eur. Respir. J., October 1, 2002; 20(4): 907 - 911. [Abstract] [Full Text] [PDF] |
||||
![]() |
S A McKenzie, E Chan, I Dundas, P D Bridge, C S Pao, M Mylonopoulou, and M J R Healy Airway resistance measured by the interrupter technique: normative data for 2-10 year olds of three ethnicities Arch. Dis. Child., September 1, 2002; 87(3): 248 - 251. [Abstract] [Full Text] [PDF] |
||||
![]() |
N. Beydon, F. Amsallem, M. Bellet, M. Boule, M. Chaussain, A. Denjean, R. Matran, B. Wuyam, C. Alberti, and C. Gaultier Pre/Postbronchodilator Interrupter Resistance Values in Healthy Young Children Am. J. Respir. Crit. Care Med., May 15, 2002; 165(10): 1388 - 1394. [Abstract] [Full Text] [PDF] |
||||
![]() |
G.L. Hall, J.H. Wildhaber, M. Cernelc, and U. Frey Evaluation of the interrupter technique in healthy, unsedated infants Eur. Respir. J., December 1, 2001; 18(6): 982 - 988. [Abstract] [Full Text] [PDF] |
||||
![]() |
E Lombardi, P D Sly, G Concutelli, E Novembre, G Veneruso, G Frongia, R Bernardini, and A Vierucci Reference values of interrupter respiratory resistance in healthy preschool white children Thorax, September 1, 2001; 56(9): 691 - 695. [Abstract] [Full Text] [PDF] |
||||
![]() |
F. Child, S. Clayton, S. Davies, A.A. Fryer, P.W. Jones, and W. Lenney How should airways resistance be measured in young children: mask or mouthpiece? Eur. Respir. J., June 1, 2001; 17(6): 1244 - 1249. [Abstract] [Full Text] [PDF] |
||||
![]() |
P. J. F. M. MERKUS, J. Y. MIJNSBERGEN, W. C. J. HOP, and J. C. de JONGSTE Interrupter Resistance in Preschool Children . Measurement Characteristics and Reference Values Am. J. Respir. Crit. Care Med., May 1, 2001; 163(6): 1350 - 1355. [Abstract] [Full Text] |
||||
![]() |
K. R. Bock, P. Silver, M. Rom, and M. Sagy Reduction in Tracheal Lumen Due to Endotracheal Intubation and Its Calculated Clinical Significance Chest, August 1, 2000; 118(2): 468 - 472. [Abstract] [Full Text] [PDF] |
||||
![]() |
V. Kessler, G. Mols, H. Bernhard, C. Haberthur, and J. Guttmann Interrupter airway and tissue resistance: errors caused by valve properties and respiratory system compliance J Appl Physiol, October 1, 1999; 87(4): 1546 - 1554. [Abstract] [Full Text] [PDF] |
||||
| HOME | HELP | FEEDBACK | SUBSCRIPTIONS | ARCHIVE | SEARCH | TABLE OF CONTENTS |