Eur Respir J 1999; 14: 1156-1162
Copyright © ERS Journals Ltd 1999


Original Articles

In vitro investigations of jet-pulses for the measurement of respiratory impedance in newborns

M Schmidt, B Foitzik, O Hochmuth, and G Schmalisch

The aim of this in vitro study was to investigate the measuring range and accuracy of a miniaturized equipment for respiratory impedance (Zrs) measurements in newborns using jet-pulses. Brief flow pulses (peak flow=16 L x min(-1), width=10 ms) were generated by a jet-generator consisting of a solenoid valve and an injector, situated between pneumotachograph and outflow resistance. Serially arranged resistance-inertance-compliance (R-I-C) lung models (RM=1.3-6.4 kPa x L(-1) x s, CM=7.4-36.9 mL x kPa(-1), IM=1.5 Pa x L(-1) x s2) were used to measure the real and imaginary part of Zrs between 4 and 50 Hz and to determine R, C and I by means of the method of least squares. The median errors for R, C and I were -0.1 kPa x L(-1) x s (-2%), 2.4 mL x kPa(-1)(13%) and -0.2 Pa x L(-1) x s2 (-13%) for measurements without breathing signals and 0.11 kPa x L(-1) -s (3%), 3 mL x kPa(-1) (16%) and 0.28 Pa x L (-1) x s2 (19%) in mechanically ventilated models. During spontaneous breathing the influence of the breathing flow on Zrs was negligible. The equipment did not show any nonlinearity when different pulse amplitudes were used (Vmax=13-22 L x min(-1)). The investigations have shown that jet-pulses allow reliable measurements of respiratory impedance and have the potential to provide valuable information about lung mechanics in spontaneously breathing and mechanically ventilated newborns. The developed measuring head has a low apparatus dead space, is easy to disinfect, has standard connections and can be used as the T-piece in a ventilator circuit.


This article has been cited by other articles:


Home page
Eur Respir JHome page
G. Schmalisch, B. Foitzik, R.R. Wauer, and J. Stocks
Effect of apparatus dead space on breathing parameters in newborns: "flow-through" versus conventional techniques
Eur. Respir. J., January 1, 2001; 17(1): 108 - 114.
[Abstract] [Full Text] [PDF]




HOME HELP FEEDBACK SUBSCRIPTIONS ARCHIVE SEARCH TABLE OF CONTENTS
Copyright © 1999 by the European Respiratory Society.